share:

Florisil Cartridge Comparison on Rtx-CLPesticides2

image
GC_EV01093
ColumnRtx-CLPesticides2, 30 m, 0.32 mm ID, 0.50 µm (cat.# 11325)
SampleFlorisil Cartridge Check Standard (2,4,5-trichlorophenol) (cat.# 32017)
Injection
Inj. Vol.:1.0 µL splitless (hold 0.75 min)
Liner:3.5mm Gooseneck Splitless (cat.# 22286)
Inj. Temp.:275 °C
Oven
Oven Temp.:120 °C (hold 1 min) to 300 °C at 10 °C/min (hold 6 min)
Carrier GasHe, constant linear velocity
Linear Velocity:65 cm/sec
DetectorECD @ 330 °C
InstrumentShimadzu 2010 GC
Notestraces shown on same scale
black trace: Restek 6mL/1g Florisil SPE Tube
red trace: competitor 6mL/1g Florisil SPE Tube
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us