share:

This chromatogram may contain analytes that are synonyms of your search terms.

Method 8081B Chlorinated Pesticides on Rtx-CLPesticides2

Peaks
1.Dibromochloropropane
2.Hexachlorocyclopentadiene
3.Tetrachloro-m-xylene
4.cis-diallate
5.Hexachlorobenzene
6.trans-diallate
7.α-BHC
8.γ-BHC
9.β-BHC
10.δ-BHC
11.Heptachlor
12.Aldrin
13.Isodrin
14.Heptachlor epoxide
Peaks
15.trans-Chlordane*
16.cis-Chlordane*
17.Endosulfan I
18.4,4'-DDE
19.Dieldrin
20.Endrin/chlorobenzilate
21.4,4'-DDD
22.Endosulfan II
23.4,4'-DDT
24.Endrin aldehyde
25.Endosulfan sulfate
26.Methoxychlor
27.Endrin ketone
28.Decachlorobiphenyl
* For information regarding the nomenclature used for cis-chlordane and trans-chlordane, visit www.restek.com/chlordane-notice
image
GC_EV1145
ColumnRtx-CLPesticides2, 30 m, 0.32 mm ID, 0.25 µm (cat.# 11324)
SampleOrganochlorine Pesticide Mix AB #1 (cat.# 32291)
Pesticide Surrogate Mix, EPA 8080, 8081 (cat.# 32000)
Organochlorine Pesticide Mix C #1 (cat.# 32296)
Diluent:hexane
InjectionDirect
Liner:Drilled Uniliner
Inj. Temp.:220 °C
Oven
Oven Temp.:80 °C (hold 1 min) to 300 °C at 9 °C/min (hold 10 min)
Carrier GasHe
DetectorECD @ 310 °C
NotesSample: 0.5 - 1.0 mL solvent extracted sample exchanged into hexane
Standards and surrogates were spiked into hexane
Tube: 3 mL, 250 mg CarboPrep 90
Tube conditioning: Apply 2 mL CH2Cl2:hexane (20:80) and pass through tube
Extract cleanup: Prepare collection rack with vials, place under each tube. Add 0.5 - 1.0 mL of extracted sample to tube and collect all solutions passing through. Do not expose carbon bed to air. Add 20 mL of CH2Cl2:hexane (20:80) to tube and allow it to elute using gravity feed.
Extract concentration: Solvent exchange extract to hexane
Do not allow extracts to dry completely
Concentrate extract to 0.5 - 1.0 mL
Sample is ready for analysis.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

8081B

,

EPA method 8081

,

CLP

,

chlorinated pesticides

,

8081

,

BHC

,

Chlordane

,

Endosulfan

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us