share:

Polycyclic Aromatic Hydrocarbons on Rxi-17Sil MS (15m x 0.25mm x 0.25µm)

Peaks
1.Naphthalene
2.2-Methylnaphthalene
3.1-Methylnaphthalene
4.Acenaphthylene
5.Acenaphthene
6.Fluorene
7.Phenanthrene
8.Anthracene
9.Fluoranthene
10.Pyrene
Peaks
11.Benzo[a]anthracene
12.Chrysene
13.Triphenylene
14.Benzo[b]fluoranthene
15.Benzo[k]fluoranthene
16.Benzo[j]fluoranthene
17.Benzo[a]pyrene
18.Indeno[1,2,3-cd]pyrene
19.Dibenz[a,h]anthracene
20.Benzo[ghi]perylene
.
image
GC_EV1186
ColumnRxi-17Sil MS, 15 m, 0.25 mm ID, 0.25 µm (cat.# 14120)
Sample
Diluent:Methylene chloride
Conc.:20 ng/µL
Injection
Inj. Vol.:1 µL split (split ratio 20:1)
Liner:4 mm split Precision liner w/wool (cat.# 21022)
Inj. Temp.:275 °C
Split Vent Flow Rate:42 mL/min
Oven
Oven Temp.:80 °C (hold 1 min) to 320 °C at 15 °C/min (hold 2 min)
Carrier GasH2, constant flow
Flow Rate:2 mL/min
DetectorFID @ 340 °C
Constant Column + Constant Make-up:50 mL/min
Make-up Gas Type:N2
Data Rate:20 Hz
InstrumentAgilent/HP6890 GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

17sil

,

17silMS

,

PAH

,

svoa

,

semivolatiles

,

polynuclear aromatic hydrocarbons

,

polycyclic aromatic hydrocarbons

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us