share:

This chromatogram may contain analytes that are synonyms of your search terms.

5 ppb Volatiles in Drinking Water on Rtx-VMS by EPA Method 524.3

Peaks
1.Dichlorodifluoromethane
2.Chlorodifluoromethane
3.Chloromethane
4.Vinyl chloride
5.1,3-Butadiene
6.Bromomethane
7.Trichlorofluoromethane
8.Diethyl ether
9.1,1-Dichloroethene
10.Carbon disulfide
11.Methyl iodide
12.Allyl chloride
13.Methylene chloride
14.trans-1,2-Dichloroethene
15.Methyl acetate
16.MTBE-D3 (SS)
17.MTBE
18.tert-Butyl alcohol (TBA)
19.Diisopropyl ether (DIPE)
20.1,1-Dichloroethane
21.tert-Butyl ethyl ether (ETBE)
22.cis-1,2-Dichloroethene
23.Bromochloromethane
24.Chloroform
25.Carbon tetrachloride
26.Tetrahydrofuran
27.1,1,1-Trichloroethane
28.1,1-Dichloropropene
29.1-Chlorobutane
30.Benzene
31.tert-Amyl methyl ether (TAME)
32.1,2-Dichloroethane
33.Trichloroethene
34.1,4-Difluorobenzene
35.tert-Amyl ethyl ether (TAEE)
36.Dibromomethane
37.1,2-Dichloropropane
38.Bromodichloromethane
39.cis-1,3-Dichloropropene
40.Toluene
41.Tetrachloroethene
Peaks
42.trans-1,3-Dichloropropene
43.1,1,2-Trichloroethane
44.Ethyl methacrylate
45.Dibromochloromethane
46.1,3-Dichloropropane
47.1,2-Dibromoethane
48.Chlorobenzene-D5
49.Chlorobenzene
50.Ethylbenzene
51.1,1,1,2-Tetrachloroethane
52.m-Xylene
53.p-Xylene
54.o-Xylene
55.Styrene
56.Bromoform
57.Isopropylbenzene
58.4-Bromofluorobenzene (SS)
59.Bromobenzene
60.n-Propylbenzene
61.1,1,2,2-Tetrachloroethane
62.2-Chlorotoluene
63.1,3,5-Trimethylbenzene
64.1,2,3-Trichloropropane
65.4-Chlorotoluene
66.tert-Butylbenzene
67.Pentachloroethane
68.1,2,4-Trimethylbenzene
69.sec-Butylbenzene
70.4-Isopropyltoluene
71.1,3-Dichlorobenzene
72.1,4-Dichlorobenzene-D4
73.1,4-Dichlorobenzene
74.n-Butylbenzene
75.Hexachloroethane
76.1,2-Dichlorobenzene-D4 (SS)
77.1,2-Dichlorobenzene
78.1,2-Dibromo-3-chloropropane
79.Hexachlorobutadiene
80.1,2,4-Trichlorobenzene
81.Naphthalene
82.1,2,3-Trichlorobenzene
image
GC_EV1297
ColumnRtx-VMS, 30 m, 0.25 mm ID, 1.40 µm (cat.# 19915)
Sample524.3 internal standard/surrogate mix (cat.# 30017)
524.3 gas calibration mix (cat.# 30014)
524.3 VOA MegaMix standard (cat.# 30013)
Diluent:RO water
Conc.:5 ng/mL (5 mL sample)
Injectionpurge and trap split (split ratio 30:1)
Liner:Premium 1.0mm ID straight inlet liner (cat.# 23333.1)
Inj. Temp.:200 °C
Purge and Trap
Instrument:EST Encon Evolution
Trap Type:Vocarb 3000
Purge: 11 min, flow 40 mL/min
Dry Purge: 1 min, flow 50 mL/min
Desorb: 1 min @ 260 °C, flow 30.9 mL/min
Bake: 8 min @ 265 °C
Interface Connection:injection port
Transfer Line Temp.:150 °C
Oven
Oven Temp.:45 °C (hold 4.5 min) to 100 °C at 12 °C/min to 240 °C at 25 °C/min (hold 1.32 min)
Carrier GasHe, constant flow
Flow Rate:0.9 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
11.547–3005.4
22.935–3005.19
Transfer Line Temp.:240 °C
Analyzer Type:Quadrupole
Source Temp.:230 °C
Quad Temp.:150 °C
Electron Energy:70 eV
Solvent Delay Time:1.5 min
Tune Type:BFB
Ionization Mode:EI
InstrumentAgilent 7890A GC & 5975C MSD
AcknowledgementEST Analytical provided the Centurion robotic autosampler and Encon Evolution P&T concentrator.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

UCMR3

,

524.3

,

volatiles

,

drinking water

,

VOA

,

VMS

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us