share:

Comparison: Florisil SPE Cleanup of Blank Extract and EPA 8081 Pesticides Extract on Rtx-CLPesticides2

PeaksConc.
(ng/mL)
1.2,4,5,6-Tetrachloro-m-xylene (SS)20
2.α-BHC5
3.γ-BHC5
4.β-BHC5
5.δ-BHC5
6.Heptachlor5
7.Aldrin5
8.Heptachlor epoxide5
9.trans-Chlordane*5
10.cis-Chlordane*5
11.Endosulfan I5
PeaksConc.
(ng/mL)
12.4,4'-DDE10
13.Dieldrin10
14.Endrin10
15.4,4'-DDD10
16.Endosulfan II10
17.4,4'-DDT10
18.Endrin aldehyde10
19.Endosulfan sulfate10
20.Methoxychlor50
21.Endrin ketone10
22.Decachlorobiphenyl (SS)20
* For information regarding the nomenclature used for cis-chlordane and trans-chlordane, visit www.restek.com/chlordane-notice
image
GC_EV1299
ColumnRtx-CLPesticides2, 30 m, 0.32 mm ID, 0.25 µm (cat.# 11324)
using Rxi guard column 5 m, 0.32 mm ID (cat.# 10039)
with universal “Y” Press-Tight connector (cat.# 20406-261)
Sample2,4,5,6-Tetrachloro-m-xylene (cat.# 32027)
Decachlorobiphenyl (BZ #209) (cat.# 32029)
Organochlorine pesticide mix AB #2 (cat.# 32292)
Injection
Inj. Vol.:2 µL splitless (hold 0.75 min)
Liner:Premium 4 mm single taper w/wool (cat.# 23303.5)
Inj. Temp.:250 °C
Purge Flow:50 mL/min
Oven
Oven Temp.:110 °C (hold 0.5 min) to 320 °C at 15 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:3.5 mL/min
Detectorµ-ECD @ 330 °C
Make-up Gas Flow Rate:50 mL/min
Make-up Gas Type:N2
Data Rate:50 Hz
InstrumentAgilent/HP6890 GC
NotesA mixed standard was prepared in 1 mL hexane (see peak list for nominal concentration of each component). For cleanup, a Florisil tube (cat.# 24034) was first conditioned with 6 mL hexane. The 1 mL standard was then loaded on the tube and eluted with hexane:acetone (90:10), collecting 10 mL of eluent. The eluent was then concentrated down to 1 mL and analyzed.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

EPA 8081

,

pesticides

,

Florisil

,

SPE

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.