share:

This chromatogram may contain analytes that are synonyms of your search terms.

Comparison: Florisil SPE Cleanup of Blank Extract and EPA 8081 Pesticides Extract on Rtx-CLPesticides2

PeaksConc.
(ng/mL)
1.2,4,5,6-Tetrachloro-m-xylene (SS)20
2.α-BHC5
3.γ-BHC5
4.β-BHC5
5.δ-BHC5
6.Heptachlor5
7.Aldrin5
8.Heptachlor epoxide5
9.trans-Chlordane*5
10.cis-Chlordane*5
11.Endosulfan I5
PeaksConc.
(ng/mL)
12.4,4'-DDE10
13.Dieldrin10
14.Endrin10
15.4,4'-DDD10
16.Endosulfan II10
17.4,4'-DDT10
18.Endrin aldehyde10
19.Endosulfan sulfate10
20.Methoxychlor50
21.Endrin ketone10
22.Decachlorobiphenyl (SS)20
* For information regarding the nomenclature used for cis-chlordane and trans-chlordane, visit www.restek.com/chlordane-notice
image
GC_EV1299
ColumnRtx-CLPesticides2, 30 m, 0.32 mm ID, 0.25 µm (cat.# 11324)
using Rxi guard column 5 m, 0.32 mm ID (cat.# 10039)
with universal “Y” Press-Tight connector (cat.# 20406-261)
Sample2,4,5,6-Tetrachloro-m-xylene (cat.# 32027)
Decachlorobiphenyl (BZ #209) (cat.# 32029)
Organochlorine pesticide mix AB #2 (cat.# 32292)
Injection
Inj. Vol.:2 µL splitless (hold 0.75 min)
Liner:Premium 4 mm single taper w/wool (cat.# 23303.5)
Inj. Temp.:250 °C
Purge Flow:50 mL/min
Oven
Oven Temp.:110 °C (hold 0.5 min) to 320 °C at 15 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:3.5 mL/min
Detectorµ-ECD @ 330 °C
Make-up Gas Flow Rate:50 mL/min
Make-up Gas Type:N2
Data Rate:50 Hz
InstrumentAgilent/HP6890 GC
NotesA mixed standard was prepared in 1 mL hexane (see peak list for nominal concentration of each component). For cleanup, a Florisil tube (cat.# 24034) was first conditioned with 6 mL hexane. The 1 mL standard was then loaded on the tube and eluted with hexane:acetone (90:10), collecting 10 mL of eluent. The eluent was then concentrated down to 1 mL and analyzed.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

EPA 8081

,

pesticides

,

Florisil

,

SPE

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us