7-Minute Analysis of Organochlorine Pesticides (EPA Method 8081) on Rtx-CLPesticides and Rtx-CLPesticides2 Using Helium
Peaks | Conc. (µg/mL) | |
---|---|---|
1. | Tetrachloro-m-xylene![]() | 20 |
2. | Hexachlorobenzene![]() | 5 |
3. | α-BHC![]() | 5 |
4. | γ-BHC![]() | 5 |
5. | β-BHC![]() | 5 |
6. | δ-BHC![]() | 5 |
7. | Heptachlor![]() | 5 |
8. | Aldrin![]() | 5 |
9. | Heptachlor epoxide![]() | 5 |
10. | trans-Chlordane![]() | 5 |
11. | cis-Chlordane![]() | 5 |

GC_EV1323
Columns | Rtx-CLPesticides2 30 m, 0.32 mm ID, 0.25 µm (cat.# 11324) |
and Rtx-CLPesticides 30 m, 0.32 mm ID, 0.32 µm (cat.# 11141) | |
using Rxi guard column 5 m, 0.32 mm ID (cat.# 10039) | |
with universal “Y” Press-Tight connector (cat.# 20406-261) | |
Sample | Organochlorine pesticide mix AB #2 (cat.# 32292) |
Hexachlorobenzene (cat.# 32231) | |
2,4,5,6-Tetrachloro-m-xylene (cat.# 32027) | |
Decachlorobiphenyl (BZ #209) (cat.# 32029) | |
Diluent: | n-Hexane |
Injection | |
Inj. Vol.: | 2 µL splitless (hold 0.3 min) |
Liner: | Premium 4 mm single taper w/wool (cat.# 23303.5) |
Inj. Temp.: | 250 °C |
Purge Flow: | 50 mL/min |
Oven | |
Oven Temp.: | 120 °C to 200 °C at 45 °C/min to 230 °C at 15 °C/min to 330 °C at 30 °C/min (hold 2 min) |
Carrier Gas | He, constant flow |
Linear Velocity: | 71 cm/sec |
Detector | Micro-ECD @ 340 °C |
Make-up Gas Flow Rate: | 50 mL/min |
Make-up Gas Type: | N2 |
Data Rate: | 50 Hz |
Instrument | Agilent/HP6890 GC |