share:

Nonylphenol, Bisphenol A, 4-tert-Octylphenol, Nonylphenol Monoethoxylate, and Nonylphenol Diethoxylate on Rxi-XLB by ASTM D7065-11

PeakstR (min)Conc.
(µg/mL)
On-Column (ng)
1.Acenaphthene-d1016.2252.5
2.4-tert-Octylphenol17.7323.2
3.Nonylphenol18.9–20.016016
4.Phenanthrene-d1020.8252.5
5.4-Nonylphenol21.4323.2
6.Nonylphenol monoethoxylate22.3–24.232032
7.4-Nonylphenol monoethoxylate24.8323.2
8.Bisphenol A25.4323.2
9.Nonylphenol diethoxylate25.7–27.764064
10.Chrysene-d1229.1252.5
Peaks 3, 6, and 9 correspond to mixtures of isomers.
image
GC_EV1404
ColumnRxi-XLB, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13723)
SampleMethod 525.2 internal standard mix (cat.# 31825)
AccuStandard nonylphenol calibration standard (cat.# M-1626)
Diluent:Dichloromethane
Conc.: See peak table for individual compound concentrations and amounts on-column.
Injection
Inj. Vol.:1 µL split (split ratio 10:1)
Liner:Premium 4 mm Precision liner w/wool (cat.# 23305.5)
Inj. Temp.:250 °C
Oven
Oven Temp.:45 °C (hold 1 min) to 305 °C at 8 °C/min (hold 1 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
12.1535-5505.4
Transfer Line Temp.:300 °C
Analyzer Type:Quadrupole
Source Type:CI
Source Temp.:300 °C
Quad Temp.:150 °C
Ionization Mode:EI
InstrumentAgilent 7890B GC & 5977A MSD
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

nonylphenols

,

branched nonylphenol monoethoxylate

,

branched nonylphenol diethoxylate

,

NP

,

NP1EO

,

NP2EO

,

OP

,

BPA

,

Bisphenol A

,

surfactants

,

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us