share:

EPA and EU Phthalates on Rxi-XLB

Peaks
1.Dimethyl phthalate
2.Diethyl phthalate
3.Benzyl benzoate
4.Diisobutyl phthalate
5.Di-n-butyl phthalate
6.Bis(2-methoxyethyl) phthalate
7.Bis[4-methyl-2-pentyl] phthalate isomer 1
8.Bis[4-methyl-2-pentyl] phthalate isomer 2
9.Bis(2-ethoxyethyl) phthalate
10.Di-n-pentyl phthalate
Peaks
11.Di-n-hexyl phthalate
12.Butyl benzyl phthalate
13.Hexyl-2-ethylhexyl phthalate
14.Bis(2-butoxyethyl) phthalate
15.Bis(2-ethylhexyl) phthalate
16.Dicyclohexyl phthalate
17.Di-n-octyl phthalate
18.Diisononyl phthalate
19.Diisodecyl phthalate
20.Dinonyl phthalate
image
GC_EV1409
ColumnRxi-XLB, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13723)
SampleBenzyl benzoate (cat.# 31847)
EPA Method 8061A phthalate esters mixture (cat.# 33227)
Diluent:Methylene chloride
Conc.:50.0 µg/mL (80 μg/mL for internal standard benzyl benzoate)
Injection
Inj. Vol.:1 µL split (split ratio 20:1)
Liner:Premium 3.5 mm Precision liner w/wool (cat.# 23320.1)
Inj. Temp.:280 °C
Split Vent Flow Rate:3 mL/min
Oven
Oven Temp.:200 °C (hold 0.5 min) to 330 °C at 30 °C/min (hold 1 min)
Carrier GasHe, constant linear velocity
Linear Velocity:66.7 cm/sec, 39.5 psi, 272.3 kPa @ 200 °C
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
10.959-400-
Transfer Line Temp.:300 °C
Analyzer Type:Quadrupole
Source Temp.:280 °C
Electron Energy:70 eV
Solvent Delay Time:0.9 min
Tune Type:PFTBA
Ionization Mode:EI
InstrumentShimadzu 2010 GC & QP2010+ MS
NotesThe flow rate is 3 mL/min @ 200 °C. The MS scan interval is 0.1 sec.
AcknowledgementThe authors would like to thank Shimadzu Corporation for their consultation with the operation of the QP2010 Plus GC-MS instrument.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

phthalate

,

GC-MS

,

EPA

,

EU

,

,

Rxi-XLB

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us