SPME Arrow vs. Traditional SPME Fiber

image
GC_EV1439
ColumnRtx-VMS, 30 m, 0.25 mm ID, 1.40 µm (cat.# 19915)
Sample2-Chloroethyl vinyl ether (cat.# 30265)
8260 MegaMix calibration mix (cat.# 30633)
8260 Surrogate mix (cat.# 30073)
VOA calibration mix #1 (ketones) (cat.# 30006)
California oxygenates mix (cat.# 30465)
502.2 Calibration mix #1 (gases) (cat.# 30042)
8260 Internal standard mix (cat.# 30074)
Diluent:Water
Conc.:50 ng/mL
Injectionsplit (split ratio 10:1)
Liner:Topaz 1.8 mm ID straight/SPME inlet liner (cat.# 23280.1)
Inj. Temp.:250 °C
Oven
Oven Temp.:40 °C (hold 1 min) to 100 °C at 12 °C/min (hold 0 min) to 240 °C at 25 °C/min (hold 1.6 min)
Carrier GasHe, constant flow
Flow Rate:1.0 mL/min
Linear Velocity:44.411 cm/sec @ 40 °C
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
1035-22612.1
Transfer Line Temp.:250 °C
Analyzer Type:Quadrupole
Source Type:Extractor
Extractor Lens:3 mm ID
Source Temp.:300 °C
Quad Temp.:150 °C
Electron Energy:70.0 eV
Tune Type:manual
Ionization Mode:EI
InstrumentAgilent 7890B GC & 5977A MSD
AcknowledgementCTC Analytics AG
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us