share:

16 Priority PAHs and Methylnaphthalenes on Rxi-5Sil MS using the GC Accelerator Kit in a 120 V Oven

PeakstR (min)Conc.
(µg/mL)
1.Naphthalene-d8 (IS)3.442.0
2.Naphthalene3.451.0
3.2-Methylnaphthalene3.861.0
4.1-Methylnaphthalene3.921.0
5.2-Fluorobiphenyl (SS)4.082.0
6.Acenaphthylene4.411.0
7.Acenaphthene-d10 (IS)4.492.0
8.Acenaphthene4.511.0
9.Fluorene4.831.0
10.2,4,6-Tribromophenol (SS)4.972.0
11.Phenanthrene-d10 (IS)5.402.0
12.Phenanthrene5.411.0
13.Anthracene5.451.0
14.Fluoranthene6.151.0
15.Pyrene-d10 (SS)6.282.0
16.Pyrene6.291.0
17.p-Terphenyl-D14 (SS)6.372.0
18.Benz[a]anthracene7.121.0
19.Chrysene-d12 (IS)7.132.0
20.Chrysene7.151.0
21.Benzo[b]fluoranthene8.171.0
22.Benzo[k]fluoranthene8.211.0
23.Benzo[a]pyrene8.561.0
24.Perylene-d12 (IS)8.642.0
25.Indeno[1,2,3-cd]pyrene10.161.0
26.Dibenz[a,h]anthracene10.201.0
27.Benzo[ghi]perylene10.581.0
image
GC_EV1472
ColumnRxi-5Sil MS, 20 m, 0.15 mm ID, 0.15 µm (cat.# 43816)
SampleEPA Method 8310 PAH mixture (cat.# 31874)
Revised SV internal standard mix (cat.# 31886)
Revised B/N surrogate mix (cat.# 31888)
Acid surrogate mix (4/89 SOW) (cat.# 31063)
Diluent:Dichloromethane
Injection
Inj. Vol.:1.0 µL split (split ratio 20:1)
Liner:Topaz 4 mm single taper w/wool (cat.# 23303)
Inj. Temp.:275 °C
Oven
Oven Temp.:60 °C (hold 0.7 min) to 285 °C at 39.8 °C/min to 305 °C at 4.3 °C/min to 320 °C at 28.5 °C/min (hold 3.5 min)
Carrier GasHe, constant flow
Flow Rate:1.0 mL/min
DetectorMS
Mode:SIM
SIM Program:
GroupStart Time
(min)
Ion(s) (m/z)Dwell (ms)
13.29102, 108, 128, 136 25
23.7185, 115, 142.1, 172.1 20
34.2876, 82, 152.1, 153.1, 164.1 20
44.7182.40, 142.90, 166.1, 329.8 25
55.2489, 94, 178.1, 188.1 15
65.88101, 106.1, 122.1, 202.1, 212.1, 244.1 20
76.83101, 120.1, 226.1, 228.1, 240.2 10
87.77126, 252.1 25
98.44126, 132.1, 252.1, 264.2 25
109.57138, 139, 276.1, 278.1 25
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Type:Extractor
Extractor Lens:9 mm ID
Source Temp.:330 °C
Quad Temp.:180 °C
Solvent Delay Time:1 min
Tune Type:DFTPP
Ionization Mode:EI
InstrumentAgilent 7890B GC & 5977A MSD
NotesFast SIM analysis of 16 priority PAHs plus the methylnaphthalenes in a 120 V oven equipped with the GC Accelerator kit (cat.# 23849) (injected 1 µg/mL = 0.05 ng on-column).
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Fast GC

,

PAH

,

Benzo fluoranthene

,

benzo[a]pyrene

,

benzo[b]fluoranthene

,

methylnaphthalene

,

SIM PAH

,

PAH SIM

,

GC Accelerator

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.