Example of Vial Cap or Injection Port Septum Bleed

GC_EX00980
Column | Rtx-5MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12623) |
Sample | Methylene chloride blank |
Injection | |
Inj. Vol.: | 1.0 µL split (split ratio 10:1) |
Liner: | 4.0 mm ID straight inlet liner w/wool (standard deactivation) (cat.# 20782) |
Inj. Temp.: | 240 °C |
Oven | |
Oven Temp.: | 70 °C (hold 1 min) to 290 °C at 20 °C/min (hold 13 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.2 mL/min |
Detector | FID @ 250 °C |
RELATED CHROMATOGRAM LIBRARY SEARCHES
septum bleed
,septa bleed
,bleed contamination
,vial septum
,interfering peaks
,troubleshooting
,Coring
,bleed
,contamination