Septum Bleed Causing Significant Interference with Target Analytes

Peaks
1.Amphetamine
2.Methamphetamine
Peaks
3.MDA
4.MDMA
5.MDEA
image
GC_EX00981
ColumnRtx-5MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12623)
SampleAmphetamine, methamphetamine, MDA, MDMA, and MDEA extracted from methanol and HFAA derivatized
Conc.:100 µg/mL each
Injection
Inj. Vol.:1 µL splitless (hold 0.5 min)
Liner:3.5 mm Custom splitless inlet liner w/ standard deactivated wool
Inj. Temp.:220 °C
Oven
Oven Temp.:70 °C (hold 1 min) to 290 °C at 15 °C/min (hold 4 min)
Carrier GasHe, constant flow
Flow Rate:1.25 mL/min
DetectorMS
Mode:Scan
Transfer Line Temp.:280 °C
Ionization Mode:EI
Scan Range:43-450 amu
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

septum

,

septa

,

vial cap

,

injection septum

,

bleed

,

septum bleed

,

septa bleed

,

contamination

,

troubleshooting