Pro EZGC Modeling of Minnesota Dept. of Ag. List 1 Pesticides on Rtx-440

PeakstR (min)Conc.
(µg/mL)
1.EPTC 8.462100
2.Propachlor15.768100
3.Ethalfluralin16.665100
4.Trifluralin17.074100
5.Desisopropyl-atrazine17.278100
6.Desethyl-atrazine17.503100
7.Phorate17.776100
8.Prometon19.485100
9.Simazine19.702100
10.Terbufos19.767100
11.Atrazine19.810100
PeakstR (min)Conc.
(µg/mL)
12.Propazine19.880100
13.Fonofos20.079100
14.Triallate20.700100
15.Dimethenamid21.624100
16.Acetochlor21.741100
17.Alachlor21.968100
18.Metribuzin22.172100
19.Metolachlor22.810100
20.Chlorpyrifos22.861100
21.Cyanazine23.286100
22.Pendimethalin23.532100
image
GC_EX1135
ColumnRtx-440, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12923)
SampleMinnesota Ag List 1 pesticide kit (cat.# 32408)
Diluent:Acetone
Conc.:100 ppm
Injection
Inj. Vol.:1 µL split (split ratio 25:1)
Liner:Premium 4 mm Precision liner w/wool (cat.# 23305.1)
Inj. Temp.:300 °C
Oven
Oven Temp.:100 °C (hold 0.5 min) to 175 °C at 4 °C/min to 250 °C at 14.5 °C/min
Carrier GasHe, constant flow
Flow Rate:2.0 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
11.640-3505
Transfer Line Temp.:300 °C
Analyzer Type:Quadrupole
Source Type:Inert
Drawout Plate:6 mm ID
Source Temp.:250 °C
Quad Temp.:180 °C
Electron Energy:70 eV
Solvent Delay Time:1.6 min
Tune Type:PFTBA
Ionization Mode:EI
InstrumentAgilent 7890A GC & 5975C MSD
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Pro EZGC

,

Chromatogram Modeler

,

Method Development

,

Herbacides

,

Minnesota Department of Agriculture List 1

,

Pesticides

,

Rtx-440

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us