share:

Butter Triglycerides on Rtx-65TG

image
GC_FF00137
ColumnRtx-65TG, 30 m, 0.25 mm ID, 0.10 µm (cat.# 17008)
SampleButterfat
Diluent:Isooctane
Conc.: 1%
Injection
Inj. Vol.:0.2 µL cold on-column
Oven
Oven Temp.:80 °C (hold 1 min) to 240 °C at 30 °C/min to 360 °C at 4 °C/min (hold 5 min)
Carrier GasH2, constant flow
Linear Velocity:70 cm/sec
DetectorFID @ 380 °C
NotesSPI injector: high performance capillary insert 60 °C,
300 °C/min to 400 °C (hold 5 min)

FID sensitivity: 16 x 10-11 AFS