FAMEs (Saw Palmetto) on Stabilwax

Peaks
1.Caproic acid
2.Caprylic acid
3.Pelargonic acid
4.Capric acid
5.Lauric acid
6.Myristic acid
7.Palmitic acid
8.Margaric acid
9.Stearic acid
10.Oleic acid
11.Linoleic acid
12.Linolenic acid
image
GC_FF00367
ColumnStabilwax, 50 m, 0.25 mm ID, 0.25 µm (cat.# 10626-105)
Injection
Inj. Vol.:1.0 µL split
Inj. Temp.:230 °C
Split Vent Flow Rate:37.5 mL/min
Oven
Oven Temp.:110 °C (hold 1 min) to 240 °C at 8 °C/min (hold 25 min)
Carrier GasH2, constant flow
Flow Rate:2.5 mL/min
DetectorFID @ 250 °C
NotesSeptum purge: 3 mL/min