share:

This chromatogram may contain analytes that are synonyms of your search terms.

Alcoholic Beverages (Acids and Esters): Caproic Acid and Ethyl Laurate on Stabilwax-DA

Peaks
1.Caproic acid
2.Ethyl laurate
image
GC_FF00525
ColumnStabilwax-DA, 30 m, 0.18 mm ID, 0.18 µm (cat.# 550752)
Sample
Diluent:ethyl acetate
Conc.:100 ppm
Injection
Inj. Vol.:1.0 µL splitless (hold 0.5 min)
Liner:4 mm splitless liner with wool (cat.# 20814-202.1)
Inj. Temp.:240 °C
Oven
Oven Temp.:80 °C to 230 °C at 5 °C/min
Carrier GasH2, constant pressure (28 psi, 193.1 kPa)
Temp.:240 °C
DetectorFID
Make-up Gas Type:N2
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us