share:

Acrylamide (Reference Standard) on Stabilwax

Peaks
1.Acrylamide
image
GC_FF00642
ColumnStabilwax, 15 m, 0.53 mm ID, 0.50 µm (cat.# 10637)
SampleAcrylamide standard
Diluent:Water
Conc.:25 µg/mL
Injection
Inj. Vol.:1 µL splitless (hold 0.5 min)
Liner:Auto SYS splitless w/wool (2 mm ID) (cat.# 20830)
Inj. Temp.:260 °C
Oven
Oven Temp.:100 °C (hold 0.5 min) to 200 °C at 15 °C/min
Carrier GasHe, constant pressure
Linear Velocity:62 cm/sec @ 100 °C
DetectorFID @ 260 °C
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Acrylamide

,

Stabilwax®

,

stabilwax

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.