share:

This chromatogram may contain analytes that are synonyms of your search terms.

Underivatized Cholesterol on Rxi-5ms (Temp. program)

Peaks
1.5-α-Cholestane (IS)
Peaks
2.Cholesterol
image
GC_FF00881
ColumnRxi -5ms, 15 m, 0.25 mm ID, 0.25 µm (cat.# 13420)
Sample1,000 μg/mL cholesterol in DMF
1,000 μg/mL 5-α-cholestane in hexane
25 ng cholesterol, 150 ng 5-α-cholestane on column
Injection
Inj. Vol.:1 µL split (split ratio 20:1)
Liner:Splitless taper (4 mm) w/wool (cat.# 22405)
Inj. Temp.:250 °C
Oven
Oven Temp.:200 °C (hold 1 min) to 330 °C at 20 °C/min (hold 7.5 min)
Carrier GasHe, constant pressure (9.7 psi, 66.9 kPa)
Linear Velocity:24 cm/sec @ 200 °C
DetectorFID @ 340 °C
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Cholesterol

,

Rxi®-5ms

,

Underivatized

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us