share:

This chromatogram may contain analytes that are synonyms of your search terms.

Egg Pasta Sterols & Glycerides on Rtx-65TG

Peaks
1.Squalene (IS)
2.Cholesterol
3.β-Sitosterol
4.Diglycerides
5.Triglycerides
image
GC_FF01046
ColumnRtx-65TG, 30 m, 0.25 mm ID, 0.10 µm (cat.# 17008)
SampleFat extract from egg pasta in diethyl ether solution with 3,000 ppm squalene (IS)
Conc.:50 µg/mL
Injection
Inj. Vol.:0.5 µL PTV split (split ratio 80:1)
Inlet Temp. Program:70 °C (hold 12 min) to 370 °C at 99 °C/min (hold 5 min)
Oven
Oven Temp.:220 °C (hold 2 min) to 360 °C at 5 °C/min (hold 5 min)
Carrier GasH2, constant flow
Flow Rate:1.5 mL/min
DetectorFID @ 370 °C
AcknowledgementDaniele Naviglio, Fabiana Pizzolongo; Dipartimento di Scienza degli Alimenti – Università degli Studi di Napoli “Federico II” – Via Università, 100 - 80055 Portici (NA) – Italia
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us