share:

Analysis of Melamine and Related Compounds in Infant Formula on Rxi-5Sil MS

PeakstR (min)Quant. IonQual. IonQual.IonQual. Ion
1.Cyanuric acid10.258345330346347
2.Ammelide11.073344329345330
3.Ammeline11.760328343329344
4.Melamine12.328327342328343
5.Benzoguanamine14.538316331332330
image
GC_FF01095
ColumnRxi-5Sil MS w/5 m Integra-Guard column, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623-124)
SampleInfant formula fortified at 1 μg/g with melamine and related analogs stock standard (cat.# 33253)
Benzoguanamine (cat.# 33251)
Conc.:0.01 µg/mL as tri-TMS derivatives
Injection
Inj. Vol.:1.0 µL splitless (hold 1 min)
Liner:3.5 mm Gooseneck Splitless w/Wool (cat.# 22286-200.1)
Inj. Temp.:280 °C
Oven
Oven Temp.:75 °C to 320 °C at 15 °C/min (hold 4 min)
Carrier GasHe, constant flow
Flow Rate:1 mL/min
DetectorMS
Mode:SIM
Transfer Line Temp.:290 °C
Ionization Mode:EI
InstrumentShimadzu 2010 GC & QP2010+ MS
Notesall method ions in table, only quantification ions were plotted
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

infant formula

,

melamine

,

cyanuric

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us