share:

This chromatogram may contain analytes that are synonyms of your search terms.

Carfentrazone Ethyl in Dandelion Root Powder (Dietary Supplement) on Rxi-5Sil MS

image
GC_FF1172
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
Sampledandelion root powder spiked with 46 pesticides
QuEChERS Internal Standard Mix for GC/MS Analysis (cat.# 33267)
anthracene for quality control (cat.# 33264)
Diluent:toluene
Conc.:80 pg/µL
Injection
Inj. Vol.:1 µL splitless (hold 1.5 min)
Liner:5mm Splitless with wool (cat.# 22975-200.1)
Inj. Temp.:250 °C
Oven
Oven Temp.:90 °C (hold 1.5 min) to 340 °C at 8 °C/min
Carrier GasHe, constant flow
Flow Rate:1.5 mL/min
DetectorMS
Mode:
Analyzer Type:TOF
Source Temp.:225 °C
Electron Energy:70 eV
Ionization Mode:EI
Acquisition Range:45-550 amu
Spectral Acquisition Rate:5 spectra/sec.
InstrumentLECO Pegasus 4D GCxGC-TOFMS
NotesPurge valve time: 1.5 min.
Detector mode: full mass range

Sample Preparation:
Wetting: 1 g of dandelion root powder combined with 9 mL of water, shaken well, fortified with pesticides and QuEChERS Internal Standard Mix for GC/MS Analysis (cat.# 33267), soak for 2 hours.

Extraction: 10 mL acetonitrile added then addition of Q-sep Q110 (cat.# 26213), centrifuge with Q-sep 3000 Centrifuge (cat.# 26230).

cSPE: add control standard anthracene (cat.# 33264) to 1 mL extract, add magnesium sulfate to cartridge, use 6 mL Combo SPE Cartridge containing 500 mg CarboPrep 90/500 mg PSA (cat.# 26194), concentrate via evaporation.

RELATED CHROMATOGRAM LIBRARY SEARCHES

tof

,

time of flight

,

EN 15562

,

wetting

,

dry

,

CarboPrep 90

,

anthracene

,

magnesium sulfate

,

GCB

,

GC-TOFMS

,

. Dispersive SPE

,

botanical

,

natural products

,

nutraceuticals

,

pesticide residues

,

GMP

,

herbal

,

QuEChERS

,

SPE

,

Food Drug and Cosmetic Act

,

DSHEA

,

PSA

,

LECO

,

Rxi-5SilMS

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us