share:

This chromatogram may contain analytes that are synonyms of your search terms.

Sage QuEChERS Extract Cleaned with Cartridge SPE (Zoom of GCxGC-TOFMS Contour Plot)

image
GC_FF1192
ColumnRxi-5Sil MS 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
Rtx-200 1.5 m, 0.18 mm ID, 0.20 µm (cat.# 45001)
Sample
Diluent:Toluene
Injection
Inj. Vol.:1 µL splitless (hold 1 min)
Liner:Gooseneck Splitless (4mm) w/Wool (cat.# 22405)
Inj. Temp.:250 °C
Purge Flow:40 mL/min
Oven
Oven Temp.:Rxi-5Sil MS: 80 °C (hold 1 min) to 310 °C at 4 °C/min (hold 1.5 min)
Rtx-200: 90 °C (hold 1 min) to 320 °C at 4 °C/min (hold 1.5 min)
Carrier GasHe, constant flow
Flow Rate:1.8 mL/min
Modulation
Modulator Temp. Offset:25 °C
Second Dimension Separation Time:4 sec
Hot Pulse Time:1.2 sec
Cool Time between Stages:0.8 sec
DetectorTOFMS
Transfer Line Temp.:290 °C
Analyzer Type:TOF
Source Temp.:225 °C
Electron Energy:70 eV
Mass Defect:-20 mu/100 u
Solvent Delay Time:4 min
Ionization Mode:EI
Acquisition Range:45 to 550 amu
Spectral Acquisition Rate:100 spectra/sec
InstrumentLECO Pegasus 4D GCxGC-TOFMS
NotesSee application note GNAN1338 for extraction and cleanup details. A 1.5 m length of the Rtx-200 column was trimmed from a 10 m column. Columns were connected with a Universal Press-Tight Connector (cat.# 20429). See chromatogram GC_FF1190 for full scale view.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

GCxGC

,

pesticides

,

dietary supplements

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us