share:

This chromatogram may contain analytes that are synonyms of your search terms.

Comparison of Spinach Extracts (TICs) Processed with Different QuEChERS dSPE Sorbents on Rxi-5Sil MS

image
GC_FF1200
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
SampleSpinach processed with QuEChERS sample preparation
QuEChERS Internal Standard Mix for GC/MS Analysis (cat.# 33267)
anthracene for quality control (cat.# 33264)
Diluent:acetonitrile
Injection
Inj. Vol.:1 µL splitless (hold 1.5 min)
Liner:Gooseneck Splitless (5mm) (cat.# 22973-200.1)
Inj. Temp.:250 °C
Oven
Oven Temp.:90 °C (hold 1.5 min) to 340 °C at 10 °C/min
Carrier GasHe, constant flow
Flow Rate:2 mL/min
DetectorMS
Mode:
Analyzer Type:TOF
Source Temp.:225 °C
Electron Energy:70 eV
Ionization Mode:EI
Acquisition Range:45-550 amu
Spectral Acquisition Rate:5 spectra/sec
InstrumentLECO Pegasus 4D GCxGC-TOFMS
NotesSample Preparation:
Sample: 10 g of homogenized spinach, fortified with pesticides and QuEChERS Internal Standard Mix for GC/MS Analysis (cat.# 33267).

Extraction: 10 mL acetonitrile added, 1 minute shake, then addition of Q-sep Q110 (cat.# 26235), 1 minute shake, centrifuge with Q-sep 3000 Centrifuge (cat.# 26230).

dSPE: Cleanup procedure according to EN 15662 method, add control standard anthracene (cat.# 33264) to 1 mL extract, add this to one of two different Q-sep dSPE tubes (either cat.# 26218 or cat.# 26219 ), shake, centrifuge.
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us