This chromatogram may contain analytes that are synonyms of your search terms.
Extracted Ion Chromatograms and Spectra for Incurred Pesticides in Strawberries on Rxi-5Sil MS

GC_FF1201
Column | Rxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623) |
Sample | Strawberries processed with QuEChERS sample preparation |
QuEChERS Internal Standard Mix for GC/MS Analysis (cat.# 33267) | |
anthracene for quality control (cat.# 33264) | |
Diluent: | acetonitrile |
Conc.: | captan incurred in store bought strawberries at 1,400 ppb and thiabendazole incurred in Amish strawberries at about 10 ppb |
Injection | |
Inj. Vol.: | 1 µL splitless (hold 1.5 min) |
Liner: | Gooseneck Splitless (5mm) (cat.# 22973-200.1) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 90 °C (hold 1.5 min) to 340 °C at 10 °C/min |
Carrier Gas | He, constant flow |
Flow Rate: | 2 mL/min |
Detector | MS |
Mode: | |
Analyzer Type: | TOF |
Source Temp.: | 225 °C |
Electron Energy: | 70 eV |
Ionization Mode: | EI |
Acquisition Range: | 45-550 amu |
Spectral Acquisition Rate: | 5 spectra/sec |
Instrument | LECO Pegasus 4D GCxGC-TOFMS |
Notes | Sample Preparation: Sample: 10 g of homogenized strawberries, either Amish or store bought, and QuEChERS Internal Standard Mix for GC/MS Analysis (cat.# 33267). Extraction: 10 mL acetonitrile added, 1 minute shake, then addition of Q-sep Q110 (cat.# 26235), 1 minute shake, centrifuge with Q-sep 3000 Centrifuge (cat.# 26230). dSPE: Cleanup procedure according to EN 15662 method, add control standard anthracene (cat.# 33264) to 1 mL extract, add this to Q-sep Q210 dSPE tubes (cat.# 26215), shake, centrifuge. Analysis: Extracted ion chromatograms of m/z 79 and 149 for captan and m/z 174 and 201 for thiabendazole. |
RELATED CHROMATOGRAM LIBRARY SEARCHES
Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.