share:

This chromatogram may contain analytes that are synonyms of your search terms.

QuEChERS Extract of Cannabis on Rxi-17Sil MS and Rxi-5ms by GCxGC-TOFMS (Zoomed in on cannabinoids region)

image
GC_FF1209
ColumnRxi-17Sil MS 30 m, 0.25 mm ID, 0.25 µm (cat.# 14123)
Rxi-5ms 0.9 m, 0.18 mm ID, 0.18 µm,
Injection
Inj. Vol.:1 µL split (split ratio 20:1)
Liner:Premium 4 mm Precision Inlet Liner w/Wool (cat.# 23305.1)
Inj. Temp.:275 °C
Oven
Oven Temp.:Rxi-17Sil MS: 40 °C (hold 1 min) to 320 °C at 5 °C/min (hold 3 min)
Rxi-5ms: 45 °C (hold 1 min) to 325 °C at 5 °C/min (hold 3 min)
Carrier GasHe, corrected constant flow (2 mL/min)
Modulation
Modulator Temp. Offset:20 °C
Second Dimension Separation Time:2 sec
Hot Pulse Time:0.6 sec
Cool Time between Stages:0.4 sec
DetectorTOFMS
Transfer Line Temp.:310 °C
Analyzer Type:TOF
Source Temp.:250 °C
Electron Energy:70 eV
Mass Defect:-20 mu/100 u
Solvent Delay Time:4.5 min
Tune Type:PFTBA
Ionization Mode:EI
Acquisition Range:45-500 amu
Spectral Acquisition Rate:200 spectra/sec
InstrumentLECO Pegasus 4D GCxGC-TOFMS
NotesSample Preparation:
Sample: Up to 2 g of finely ground cannabis in a 50 mL polypropylene centrifuge tube.

Extraction: Add 10 mL acetonitrile and 10 mL of organic-free water, then shake thoroughly. Soak 1 hour, then aggressively vortex 30 minutes. Add Q-sep packet (cat.# 26235), shake 1 minute, then centrifuge 5 minutes at 3,000 g with Q-sep 3000 centrifuge (cat.# 26230).
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Medical marijuana

,

Cannabis

,

Cannabinoids

,

GCxGC

,

GCxGC-TOFMS

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us