This chromatogram may contain analytes that are synonyms of your search terms.
GCxGC-TOFMS Chromatogram of QuEChERS Extract of Red Bell Pepper Showing Incurred Pesticides Using Rxi-5Sil MS and Rtx-200
Peaks | |
---|---|
1. | Endosulfan I![]() |
2. | Endosulfan II![]() |
3. | Endosulfan sulfate![]() |

GC_FF1220
Column | Rxi-5Sil MS 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623) |
Rtx-200 1.5 m, 0.18 mm ID, 0.20 µm (cat.# 45001) | |
Sample | Pesticides in red bell pepper extract |
QuEChERS internal standard mix for GC-MS analysis (cat.# 33267) | |
Diluent: | Acetonitrile |
Injection | |
Inj. Vol.: | 1 µL splitless (hold 1 min) |
Liner: | Gooseneck splitless (5 mm) w/deactivated wool (cat.# 22973-200.1) |
Inj. Temp.: | 250 °C |
Purge Flow: | 40 mL/min |
Oven | |
Oven Temp.: | Rxi-5Sil MS: 90 °C (hold 1 min) to 310 °C at 4 °C/min (hold 2 min) Rtx-200: 100 °C (hold 1 min) to 320 °C at 4 °C/min (hold 2 min) |
Carrier Gas | He, corrected constant flow (1.8 mL/min) |
Modulation | |
Modulator Temp. Offset: | 25 °C |
Second Dimension Separation Time: | 4 sec |
Hot Pulse Time: | 1.2 sec |
Cool Time between Stages: | 0.8 sec |
Detector | TOFMS |
Transfer Line Temp.: | 290 °C |
Analyzer Type: | TOF |
Source Temp.: | 225 °C |
Electron Energy: | 70 eV |
Mass Defect: | -20 mu/100 u |
Solvent Delay Time: | 5 min |
Tune Type: | PFTBA |
Ionization Mode: | EI |
Acquisition Range: | 45-550 amu |
Spectral Acquisition Rate: | 100 spectra/sec |
Instrument | LECO Pegasus 4D GCxGC-TOFMS |
Notes | Rtx-200 (cat.# 45001) is a 10 m column. A 1.5 m section was cut off and used as the second dimension column. Internal Standard: 100 µL of QuEChERS internal standard mix for GC-MS analysis (cat.# 33267) and 100 µL of QuEChERS internal standard mix for LC-MS/MS analysis (cat.# 33261) were added to each 10 g sample. These internal standard mixes require no dilutions (“snap-and-shoot”) and contain compounds specified in the EN 15662 QuEChERS method. Extraction: 10 mL acetonitrile was added to a 10 gram sample of homogenized red bell pepper. After a 1 minute shake, Q-sep European method extraction salts (cat.# 26236) were added followed by another 1 minute shake. The sample was then centrifuged with Q-sep 3000 centrifuge (cat.# 26230). dSPE: Add 1 mL extract to QuEChERS dSPE cleanup tube containing 150 mg MgSO4, 25 mg PSA, and 25 mg C18 (cat.# 26216). Shake for 30 seconds and centrifuge with Q-sep 3000 centrifuge (cat.# 26230). |
RELATED CHROMATOGRAM LIBRARY SEARCHES
Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.