Application-Specific QC Test on Rxi®-PAH (60 m x 0.25 mm x 0.10 µm)

Peaks
1.Triphenylene
2.Chrysene
3.Benzo[b]fluoranthene
4.Benzo[k]fluoranthene
5.Benzo[j]fluoranthene
image
GC_FF1236
ColumnRxi®-PAH, 60 m, 0.25 mm ID, 0.10 µm (cat.# 49317)
SampleQC test mix
Diluent:Toluene
Conc.:25-250 µg/mL
Injection
Inj. Vol.:1 µL split
Liner:4 mm ID Precision® low pressure drop liner w/deact. wool (cat.# 21032)
Inj. Temp.:275 °C
Split Vent Flow Rate:211 mL/min
Oven
Oven Temp.:120 °C (hold 23 min) to 240 °C at 8 °C/min (hold 18 min) to 285 °C at 4 °C/min to 350 °C at 25 °C/min (hold 10.15 min)
Carrier GasH2, constant flow
Linear Velocity:38 cm/sec
DetectorFID @ 350 °C
InstrumentAgilent/HP6890 GC
Download PDF