Application-Specific QC Test on Rxi®-PAH (30 m x 0.25 mm x 0.10 µm)
Peaks | |
---|---|
1. | Triphenylene![]() |
2. | Chrysene![]() |
3. | Benzo[b]fluoranthene![]() |
4. | Benzo[k]fluoranthene![]() |
5. | Benzo[j]fluoranthene![]() |

GC_FF1239
Column | Rxi®-PAH, 30 m, 0.25 mm ID, 0.10 µm (cat.# 49318) |
Sample | QC test mix |
Diluent: | Toluene |
Conc.: | 25-250 µg/mL |
Injection | |
Inj. Vol.: | 1 µL split |
Liner: | 4 mm ID Precision® low pressure drop liner w/deact. wool (cat.# 21032) |
Inj. Temp.: | 275 °C |
Split Vent Flow Rate: | 164 mL/min |
Oven | |
Oven Temp.: | 120 °C (hold 11 min) to 240 °C at 16 °C/min (hold 9 min) to 285 °C at 8 °C/min to 350 °C at 25 °C/min (hold 9.78 min) |
Carrier Gas | H2, constant flow |
Linear Velocity: | 38 cm/sec |
Detector | FID @ 350 °C |
Instrument | Agilent/HP6890 GC |