share:

GCxGC-TOFMS Analysis of Multi-Pesticide Standard Using Rxi-5Sil MS and Rtx-200

Peaks
1.Methamidophos
2.Dichlorvos
3.1,2,3,5-Tetrachlorobenzene
4.Mevinphos
5.Acephate
6.Pentachlorobenzene
7.o-Phenylphenol
8.Tetrachloronitrobenzene (Tecnazene)
9.Omethoate
10.2,3,5,6-Tetrachloroaniline
11.Alpha-BHC
12.Hexachlorobenzene
13.Pentachloroanisole
14.Dimethoate
15.Beta-BHC
16.Pentachloronitrobenzene
17.Pentachlorobenzonitrile
18.Gamma-BHC
19.Chlorothalonil
20.Anthracene
21.Diazinon
22.delta-BHC
23.Pentachloroaniline
24.Vinclozolin
25.Carbaryl
26.Metalaxyl
27.Pentachlorothioanisole
28.Pirimiphos methyl
29.Methiocarb
30.Dichlofluanid
31.Malathion
32.Chlorpyrifos
33.Fenthion
34.DCPA methyl ester (Chlorthal-dimethyl)
35.Parathion
36.Cyprodinil
Peaks
37.Heptachlor epoxide
38.Thiabendazole
39.Captan
40.Folpet
41.Procymidone
42.Endosulfan I
43.Imazalil
44.4,4'-DDE
45.Dieldrin
46.Myclobutanil
47.Endrin
48.Endosulfan II
49.Oxadixyl
50.4,4'-DDD
51.2,4'-DDT
52.Carfentrazone ethyl
53.Endosulfan sulfate
54.Fenhexamid
55.4,4'-DDT
56.Propargite
57.Piperonyl butoxide
58.Iprodione
59.Bifenthrin
60.Dicofol
61.Fenpropathrin
62.Phosalone
63.Azinphos-methyl
64.cis-Permethrin
65.Coumaphos
66.trans-permethrin
67.Cypermethrin
68.Pyraclostrobin
69.Fluvalinate
70.Difenoconazole
71.Deltamethrin
72.Azoxystrobin
image
GC_FF1241
ColumnRxi-5Sil MS 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
Rtx-200 1.3 m, 0.25 mm ID, 0.25 µm (cat.# 15124)
SampleCustom pesticide standard
Diluent:Acetonitrile
Conc.:500 pg/µL
Injection
Inj. Vol.:1.0 µL splitless (hold 1.0 min)
Liner:Premium 4 mm single taper w/wool (cat.# 23303.5)
Inj. Temp.:250 °C
Oven
Oven Temp.:Rxi-5Sil MS: 90 °C (hold 1.0 min) to 310 °C at 5 °C/min (hold 2.0 min)
Rtx-200: 95 °C (hold 1.0 min) to 315 °C at 5 °C/min (hold 2.0 min)
Carrier GasHe, corrected constant flow (2 mL/min)
Modulation
Modulator Temp. Offset:+20 °C
Second Dimension Separation Time:3 sec
Hot Pulse Time:0.9 sec
Cool Time between Stages:0.6 sec
DetectorMS
Mode:
Transfer Line Temp.:300 °C
Analyzer Type:TOF
Source Temp.:225 °C
Electron Energy:70 eV
Mass Defect:-20 mu/100 u
Ionization Mode:EI
Acquisition Range:45 to 550 amu
Spectral Acquisition Rate:100 spectra/sec
InstrumentLECO Pegasus 4D GCxGC-TOFMS
NotesRtx-200 (cat.# 15124) is a 2 m column. A 1.3 m section was cut off and used as the second dimension column.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Organonitrogen

,

Organophosphorous

,

Tobacco

,

pesticides

,

GCxGC

,

OCP

,

Organochlorine

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us