share:

This chromatogram may contain analytes that are synonyms of your search terms.

Benz[a]anthracene and Chrysene in a QuEChERS Extract of Mate Tea Separated from Potential Interfering PAHs on Rxi-PAH

PeakstR (sec)
1.Benz[a]anthracene1,028.4
2.Cyclopenta[cd]pyrene1,044.0
3.Triphenylene1,050.0
4.Chrysene1,054.8
image
GC_FF1244
ColumnRxi-PAH, 60 m, 0.25 mm ID, 0.10 µm (cat.# 49317)
Injection
Inj. Vol.:2.5 µL splitless (hold 1 min)
Liner:Premium 4 mm single taper w/wool (cat.# 23303.5)
Inj. Temp.:275 °C
Purge Flow:40 mL/min
Oven
Oven Temp.:80 °C (hold 1 min) to 210 °C at 40 °C/min to 260 °C at 3 °C/min to 350 °C at 11.5 °C/min (hold 6.25 min)
Carrier GasH2, constant flow
Flow Rate:2.4 mL/min
DetectorTOFMS
Transfer Line Temp.:320 °C
Analyzer Type:TOF
Source Temp.:300 °C
Electron Energy:70 eV
Mass Defect:0 mu/100 u
Solvent Delay Time:3.67 min
Tune Type:PFTBA
Ionization Mode:EI
Acquisition Range:45-550 amu
Spectral Acquisition Rate:5 spectra/sec
InstrumentLECO Pegasus 4D GCxGC-TOFMS
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

EFSA PAH4

,

PAHs

,

Chrysene

,

Triphenylene

,

Benz[a]anthracene

,

Cyclopenta[cd]pyrene

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us