share:

Benzo[b]fluoranthene and Benzo[a]pyrene in a QuEChERS Extract of Mate Tea Separated from Potential Interfering PAHs on Rxi-PAH

PeakstR (sec)
1.Benzo[b]fluoranthene1,403.6
2.Benzo[k]fluoranthene1,409.6
3.Benzo[j]fluoranthene1,413.6
4.Benzo[a]fluoranthene1,431.6
5.Benzo[e]pyrene1,478.2
6.Benzo[a]pyrene1,489.0
7.Perylene1,511.4
image
GC_FF1245
ColumnRxi-PAH, 60 m, 0.25 mm ID, 0.10 µm (cat.# 49317)
Injection
Inj. Vol.:2.5 µL splitless (hold 1 min)
Liner:Premium 4 mm single taper w/wool (cat.# 23303.5)
Inj. Temp.:275 °C
Purge Flow:40 mL/min
Oven
Oven Temp.:80 °C (hold 1 min) to 210 °C at 40 °C/min to 260 °C at 3 °C/min to 350 °C at 11.5 °C/min (hold 6.25 min)
Carrier GasH2, constant flow
Flow Rate:2.4 mL/min
DetectorTOFMS
Transfer Line Temp.:320 °C
Analyzer Type:TOF
Source Temp.:300 °C
Electron Energy:70 eV
Mass Defect:0 mu/100 u
Solvent Delay Time:3.67 min
Tune Type:PFTBA
Ionization Mode:EI
Acquisition Range:45-550 amu
Spectral Acquisition Rate:5 spectra/sec
InstrumentLECO Pegasus 4D GCxGC-TOFMS
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

PAHs

,

EFSA PAH4

,

Benzo[b]fluoranthene

,

Benzo[a]pyrene

,

Benzo[j]fluoranthene

,

Benzo[k]fluoranthene

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.