share:

This chromatogram may contain analytes that are synonyms of your search terms.

Nutritional Labeling FAMEs on Rt-2560 by AOCS Method Ce-1j-07

PeakstR (min)Conc.
(µg/mL)
Structural Nomenclature
1.Methyl caproate7.5240C6:0
2.Methyl octanoate7.8840C8:0
3.Methyl decanoate8.4840C10:0
4.Methyl undecanoate8.9320C11:0
5.Methyl dodecanoate9.5140C12:0
6.Methyl tridecanoate10.2720C13:0
7.Methyl myristate11.2740C14:0
8.Methyl myristoleate12.4820C14:1 (c9)
9.Methyl pentadecanoate12.5720C15:0
10.Methyl pentadecenoate14.1520C15:1 (C10)
11.Methyl palmitate14.2860C16:0
12.Methyl palmitoleate15.9820C16:1 (c9)
13.Methyl heptadecanoate16.5120C17:0
14.Methyl heptadecenoate18.6820C17:1 (c10)
15.Methyl stearate19.4340C18:0
16.Methyl octadecenoate21.0820C18:1 (t9)
17.Methyl oleate21.8540C18:1 (c9)
18.Methyl linolelaidate24.0920C18:2 (t9,t12)
19.Methyl linoleate26.1420C18:2 (c9,c12)
20.Methyl arachidate28.2540C20:0
21.Methyl linolenate29.9820C18:3 (c6,c9,c12)
22.Methyl eicosenoate32.0520C20:1 (c11)
23.Methyl linolenate32.4120C18:3 (c9,c12,c15)
24.Methyl heneicosanoate33.6620C21:0
25.Methyl eicosadienoate35.3320C20:2 (c11,c14)
26.Methyl behenate36.6440C22:0
27.Methyl eicosatrienoate37.4420C20:3 (c8,c11,c14)
28.Methyl erucate38.5120C22:1 (c13)
29.Methyl eicosatrienoate38.7220C20:3 (c11,c14,c17)
30.Methyl arachidonate39.1220C20:4 (c5,c8,c11,c14,c17)
31.Methyl tricosanoate39.7420C23:0
32.Methyl docosadienoate41.6420C22:2 (c13,c16)
33.Methyl eicosapentaenoate43.0720C20:5 (c5,c8,c11,c14,c17)
34.Methyl lignocerate43.1140C24:0
35.Methyl nervonate45.3320C24:1 (c15)
36.Methyl docosahexaenoate54.0220C22:6 (c4,c7,c10,c13,c16,c19)
image
GC_FF1262
ColumnRt-2560, 100 m, 0.25 mm ID, 0.20 µm (cat.# 13198)
SampleFood industry FAME mix (cat.# 35077)
Diluent:Hexane/dichloromethane
Conc.:1,000 µg/mL
Injection
Inj. Vol.:1 µL split (split ratio 20:1)
Liner:Premium 4 mm Precision liner w/wool (cat.# 23305.1)
Inj. Temp.:235 °C
Oven
Oven Temp.:180 °C (hold 32 min) to 215 °C at 20 °C/min (hold 31.25 min)
Carrier GasHe, constant flow
Flow Rate:2.0 mL/min
DetectorFID @ 325 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
Hydrogen flow:30 mL/min
Air flow:300 mL/min
Data Rate:20 Hz
InstrumentAgilent 7890A GC
NotesC4:0 Methyl butyrate (623-42-7) elutes in the solvent front.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Fats and Oils

,

AOCS Ce-1j-07

,

Nutritional Labeling

,

Trans Fats

,

Fatty Acid Profiling

,

FAMEs

,

High Resolution GLC

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us