Butter Flavored Popcorn on FAMEWAX

GC_FF1273
Column | FAMEWAX, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12497) |
Sample | Butter flavored popcorn |
Diluent: | Hexane |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 100:1) |
Liner: | Topaz 4 mm ID straight inlet liner w/wool (cat.# 23300) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 160 °C to 240 °C at 2 °C/min (hold 1 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.4 mL/min |
Detector | MS |
Mode: | Scan |
Transfer Line Temp.: | 250 °C |
Analyzer Type: | Quadrupole |
Source Type: | Inert |
Instrument | Agilent 7890A GC & 5975C MSD |
Notes | The sample was prepared using sodium methoxide. |