share:

Citronella Java Oil on Rtx-Wax

PeakstR (min)
1.D-Limonene1.809
2.Citronellal2.092
3.Linalool3.184
4.Isopulegol3.397
5.γ-Elemene3.561
6.β-Caryophyllene3.638
7.Citronellyl propionate3.917
8.α-Terpineol4.029
PeakstR (min)
9.Germacrene D4.290
10.Geranyl acetate4.509
11.β-Citronellol4.509
12.δ-Cadinene4.569
13.Nerol4.722
14.Geraniol5.006
15.α-Elemol6.417
16.Eugenol6.884
image
GC_FF1296
ColumnRtx-Wax, 30 m, 0.32 mm ID, 0.25 µm (cat.# 12424)
SampleCitronella java oil
Diluent:Acetone
Conc.:5%
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:230 °C
Oven
Oven Temp.:100 °C (hold 0.5 min) to 250 °C at 16 °C/min (hold 10 min)
Carrier GasH2, constant flow
Flow Rate:2 mL/min
DetectorFID @ 250 °C
Constant Column + Constant Make-up:52 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:400 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890A GC
NotesAll peaks were identified using the NIST MS EI spectra library (2005).
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

citronellal

,

geraniol

,

essential oils

,

citronellol

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us