share:

Rosemary Oil on Rxi-5Sil MS

PeakstR (min)
1.α-Thujene2.005
2.Tricyclene2.025
3.α-Pinene2.068
4.Camphene2.159
5.β-Myrcene2.258
6.β-Pinene2.296
7.α-Phellandrene2.409
8.α-Terpinene2.471
9.p-Cymene2.516
10.Limonene2.551
11.Eucalyptol2.594
12.γ-Terpinene2.720
13.Terpinolene2.927
PeakstR (min)
14.Linalool2.958
15.Camphor3.486
16.Borneol3.680
17.Terpinen-4-ol3.726
18.α-Terpineol3.842
19.Verbenone3.988
20.Bornyl acetate4.700
21.Eugenol5.371
22.Copaene5.713
23.Caryophyllene6.219
24.α-Caryophyllene6.587
25.γ-Muurolene6.744
26.δ-Cadinene7.188
image
GC_FF1311
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
SampleRosemary oil
Diluent:Acetone
Conc.:5%
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:100 °C to 300 °C at 11 °C/min (hold 10 min)
Carrier GasHe, constant flow
Flow Rate:1.31 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
11.0035-5005
Transfer Line Temp.:300 °C
Analyzer Type:Quadrupole
Source Type:Inert
Source Temp.:230 °C
Quad Temp.:150 °C
InstrumentAgilent 7890A GC & 5975C MSD
NotesAll peaks were identified using the NIST MS EI spectra library (2005).
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Essential oils

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us