share:

This chromatogram may contain analytes that are synonyms of your search terms.

Lavender Oil on Rxi-5Sil MS

PeakstR (min)
1.α-Thujene2.005
2.α-Pinene2.061
3.Camphene2.153
4.1-Octen-3-ol2.189
5.3-Octanone2.222
6.β-Myrcene2.254
7.β-Pinene2.285
8.Hexyl acetate2.335
9.β-Ocimene2.514
10.D-Limonene2.551
11.Eucalyptol2.591
12.γ-Terpinene2.719
13.Linalool Oxide2.806
PeakstR (min)
14.Terpinolene2.925
15.Linalool2.976
16.Camphor3.483
17.Borneol3.682
18.4-Carvomenthenol3.734
19.α-Terpineol3.843
20.Linalyl acetate4.275
21.Lavandulyl acetate4.577
22.Neryl acetate5.352
23.Geranyl acetate5.556
24.Caryophyllene6.218
25.β-Farnesene6.399
26.β-Cubebene6.845
27.α-Bisabolol8.876
image
GC_FF1313
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
SampleLavender oil
Diluent:Acetone
Conc.:5%
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:100 °C to 300 °C at 11 °C/min (hold 10 min)
Carrier GasHe, constant flow
Flow Rate:1.31 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
11.0035-5005
Transfer Line Temp.:300 °C
Analyzer Type:Quadrupole
Source Type:Inert
Source Temp.:230 °C
Quad Temp.:150 °C
InstrumentAgilent 7890A GC & 5975C MSD
NotesAll peaks were identified using the NIST MS EI spectra library (2005).
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Essential oils

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us