share:

Citronella Oil on Rxi-5Sil MS (10 m, 0.15 mm ID, 0.15 µm)

PeakstR (min)
1.α-Pinene0.513
2.D-Limonene0630
3.Linalool0.729
4.Citronellal0.841
5.Isopulegol isomer 10.852
6.Isopulegol isomer 20.872
7.α-Terpineol0.942
8.β-Citronellol0.994
9.Z-Citral1.027
10.Geraniol1.058
11.E-Citral1.094
12.Citronellyl propionate1.284
13.Eugenol1.313
14.Geranyl acetate1.358
PeakstR (min)
15.β-Elemene1.420
16.Caryophyllene1.513
17.α-Caryophyllene1.605
18.α-Amorphene1.643
19.Germacrene D1.669
20.α-Muurolene1.703
21.δ-Cadinene1.753
22.α-Cadinene1.803
23.Elemol1.829
24.Germacrene D1.911
25.γ-Eudesmol2.048
26.α-Cadinol2.073
27.α-Eudesmol2.105
28.β-Eudesmol2.112
image
GC_FF1332
ColumnRxi-5Sil MS, 10 m, 0.15 mm ID, 0.15 µm (cat.# 43815)
SampleCitronella oil
Diluent:Acetone
Conc.:1%
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:100 °C to 300 °C at 45 °C/min to 320 °C at 30 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.0 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
11.0035-50011
Transfer Line Temp.:300 °C
Analyzer Type:Quadrupole
Source Type:Inert
Source Temp.:230 °C
Quad Temp.:150 °C
InstrumentAgilent 7890A GC & 5975C MSD
NotesAll peaks were identified using the NIST MS EI spectra library (2005).
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Essential oils

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us