share:

Lavender Oil on Rxi-5Sil MS (10 m, 0.15 mm ID, 0.15 µm)

PeakstR (min)
1.α-Thujene0.499
2.α-Pinene0.513
3.Camphene0.535
4.1-Octen-3-ol0.544
5.β-Phellandrene0.559
6.β-Pinene0.567
7.Hexyl acetate0.579
8.β-Ocimene0.621
9.D-Limonene0.630
10.Eucalyptol0.640
11.γ-Terpinene0.670
12.Linalool oxide0.693
13.Terpinolene0.715
PeakstR (min)
14.Linalool0.736
15.Camphor0.855
16.Borneol0.902
17.4-Carvomenthenol0.915
18.α-Terpineol0.941
19.Linalyl acetate1.050
20.Lavandulyl acetate1.117
21.Neryl acetate1.304
22.Geranyl acetate1.354
23.Caryophyllene1.514
24.β-Farnesene1.558
25.β-Cubebene1.667
26.α-Bisabolol2.163
image
GC_FF1337
ColumnRxi-5Sil MS, 10 m, 0.15 mm ID, 0.15 µm (cat.# 43815)
SampleLavender oil
Diluent:Acetone
Conc.:1%
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:100 °C to 300 °C at 45 °C/min to 320 °C at 30 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.01 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
11.0035-50011
Transfer Line Temp.:300 °C
Analyzer Type:Quadrupole
Source Type:Inert
Source Temp.:230 °C
Quad Temp.:150 °C
InstrumentAgilent 7890A GC & 5975C MSD
NotesAll peaks were identified using the NIST MS EI spectra library (2005).
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Essential oils

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us