PeakstR (min)Conc. (µg/mL)Structural Nomenclature
1.Methyl caproate4.741400C6:0
2.Methyl octanoate4.954400C8:0
3.Methyl decanoate5.319400C10:0
4.Methyl undecanoate5.589200C11:0
5.Methyl dodecanoate5.942400C12:0
6.Methyl tridecanoate6.403200C13:0
7.Methyl myristate7.005400C14:0
8.Methyl myristoleate7.744200C14:1 (c9)
9.Methyl pentadecanoate7.790200C15:0
10.Methyl pentadecenoate8.745200C15:1 (C10)
11.Methyl palmitate8.818600C16:0
12.Methyl palmitoleate9.846200C16:1 (c9)
13.Methyl heptadecanoate10.153200C17:0
14.Methyl heptadecenoate11.469200C17:1 (c10)
15.Methyl stearate11.905400C18:0
16.Methyl octadecenoate12.903200C18:1 (t9)
17.Methyl oleate13.372400C18:1 (c9)
18.Methyl linolelaidate14.718200C18:2 (t9,t12)
19.Methyl linoleate15.963200C18:2 (c9,c12)
20.Methyl arachidate17.176400C20:0
21.Methyl linolenate18.295200C18:3 (c6,c9,c12)
22.Methyl eicosenoate19.470200C20:1 (c11)
23.Methyl linolenate19.787200C18:3 (c9,c12,c15)
24.Methyl heneicosanoate21.060200C21:0
25.Methyl eicosadienoate23.742200C20:2 (c11,c14)
26.Methyl behenate26.204400C22:0
27.Methyl eicosatrienoate27.448200C20:3 (c8,c11,c14)
28.Methyl erucate29.836200C22:1 (c13)
29.Methyl eicosatrienoate29.996200C20:3 (c11,c14,c17)
30.Methyl arachidonate30.667200C20:4 (c5,c8,c11,c14)
31.Methyl tricosanoate32.652200C23:0
32.Methyl docosadienoate34.142200C22:2 (c13,c16)
33.Methyl eicosapentaenoate34.934200C20:5 (c5,c8,c11,c14,c17)
34.Methyl lignocerate35.417400C24:0
35.Methyl nervonate36.878200C24:1 (c15)
36.Methyl docosahexaenoate42.513200C22:6 (c4,c7,c10,c13,c16,c19)
image
GC_FF1347
ColumnRt-2560, 100 m, 0.25 mm ID, 0.20 µm (cat.# 13198)
SampleFood industry FAME mix (cat.# 35077)
Diluent:Hexane
Conc.:10,000 µg/mL total concentration
Injection
Inj. Vol.:1 µL split (split ratio 20:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:180 °C (hold 32 min) to 215 °C at 20 °C/min (hold 30.25 min)
Carrier GasH2, constant flow
Flow Rate:2.5 mL/min
DetectorFID @ 250 °C
Constant Column + Constant Make-up:52 mL/min
Hydrogen flow:40 mL/min
Air flow:400 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890A GC
NotesHydrogen flow optimized to achieve effective linear velocity (https://blog.restek.com/?p=52224).
C4:0 Methyl butyrate (623-42-7) elutes in the solvent front.
Resolution of critical pair cis-11-C20:1 (#22) and cis-9,12,15-C18:3 (#23) is 2.03.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Oil

,

Fat

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us