share:

This chromatogram may contain analytes that are synonyms of your search terms.

Food Industry FAMEs on Rt-2560 by AOCS Method Ce-1h-05

PeakstR (min)Conc. (µg/mL)Structural Nomenclature
1.Methyl caproate5.830400C6:0
2.Methyl octanoate6.092400C8:0
3.Methyl decanoate6.539400C10:0
4.Methyl undecanoate6.870200C11:0
5.Methyl dodecanoate7.302400C12:0
6.Methyl tridecanoate7.866200C13:0
7.Methyl myristate8.603400C14:0
8.Methyl myristoleate9.509200C14:1 (c9)
9.Methyl pentadecanoate9.563200C15:0
10.Methyl pentadecenoate10.733200C15:1 (C10)
11.Methyl palmitate10.822600C16:0
12.Methyl palmitoleate12.080200C16:1 (c9)
13.Methyl heptadecanoate12.453200C17:0
14.Methyl heptadecenoate14.065200C17:1 (c10)
15.Methyl stearate14.597400C18:0
16.Methyl octadecenoate15.817200C18:1 (t9)
17.Methyl oleate16.393400C18:1 (c9)
18.Methyl linolelaidate18.037200C18:2 (t9,t12)
19.Methyl linoleate19.563200C18:2 (c9,c12)
20.Methyl arachidate21.048400C20:0
21.Methyl linolenate22.417200C18:3 (c6,c9,c12)
22.Methyl eicosenoate23.849200C20:1 (c11)
23.Methyl linolenate24.242200C18:3 (c9,c12,c15)
24.Methyl heneicosanoate25.794200C21:0
25.Methyl eicosadienoate29.073200C20:2 (c11,c14)
26.Methyl behenate32.095400C22:0
27.Methyl eicosatrienoate33.613200C20:3 (c8,c11,c14)
28.Methyl erucate36.531200C22:1 (c13)
29.Methyl eicosatrienoate36.727200C20:3 (c11,c14,c17)
30.Methyl arachidonate37.548200C20:4 (c5,c8,c11,c14)
31.Methyl tricosanoate40.227200C23:0
32.Methyl docosadienoate45.045200C22:2 (c13,c16)
33.Methyl eicosapentaenoate48.043200C20:5 (c5,c8,c11,c14,c17)
34.Methyl lignocerate51.086400C24:0
35.Methyl nervonate58.002200C24:1 (c15)
36.Methyl docosahexaenoate86.473200C22:6 (c4,c7,c10,c13,c16,c19)
image
GC_FF1349
ColumnRt-2560, 100 m, 0.25 mm ID, 0.20 µm (cat.# 13198)
SampleFood industry FAME mix (cat.# 35077)
Diluent:Hexane
Conc.:10,000 µg/mL total concentration
Injection
Inj. Vol.:1 µL split (split ratio 20:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:180 °C (hold 90 min)
Carrier GasH2, constant flow
Flow Rate:1.75 mL/min
DetectorFID @ 250 °C
Constant Column + Constant Make-up:52 mL/min
Hydrogen flow:40 mL/min
Air flow:400 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890A GC
NotesHydrogen flow optimized to achieve effective linear velocity (https://blog.restek.com/?p=52224).
C4:0 Methyl butyrate (623-42-7) elutes in the solvent front.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Oil

,

Fat

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us