This chromatogram may contain analytes that are synonyms of your search terms.
GCxGC-TOFMS Contour Plot of Unfortified Berry Tea

GC_FS0502
Column | Rxi-5ms 30 m, 0.25 mm ID, 0.25 µm (cat.# 13423) with Rxi guard column 1.0 m, 0.25 mm ID (cat.# 10029) |
Rtx-200 1.0 m, 0.25 mm ID, 0.25 µm (cat.# 15124) | |
Sample | Berry tea |
QuEChERS internal standard mix for GC-MS analysis (cat.# 33267) | |
Diluent: | Acetonitrile |
Injection | |
Inj. Vol.: | 1.0 µL splitless (hold 1.0 min) |
Liner: | Premium 4 mm single taper (cat.# 23302.1) |
Inj. Temp.: | 250 °C |
Purge Flow: | 40 mL/min |
Oven | |
Oven Temp.: | Rxi-5ms: 70 °C (hold 1.0 min) to 330 °C at 8.0 °C/min (hold 6.5 min) Rtx-200: 75 °C (hold 1.0 min) to 335 °C at 8.0 °C/min (hold 6.5 min) |
Carrier Gas | He, corrected constant flow (1.4 mL/min) |
Modulation | |
Modulator Temp. Offset: | 20 °C |
Second Dimension Separation Time: | 2.00 sec |
Hot Pulse Time: | 0.60 sec |
Cool Time between Stages: | 0.40 sec |
Detector | MS |
Mode: | |
Transfer Line Temp.: | 320 °C |
Analyzer Type: | TOF |
Source Temp.: | 225 °C |
Electron Energy: | 70 eV |
Mass Defect: | -20 mu/100 u |
Ionization Mode: | EI |
Acquisition Range: | 45-550 amu |
Spectral Acquisition Rate: | 100 spectra/sec |
Instrument | LECO Pegasus 4D GCxGC-TOFMS |
Notes | All column connections were made with an SGE SilTite µ-union. |
RELATED CHROMATOGRAM LIBRARY SEARCHES
Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.