share:

This chromatogram may contain analytes that are synonyms of your search terms.

Cannabis Residual Solvents at 50 ppm Cutoff

Peaks
1.Ethyl mercaptan
2.Isobutane
3.Butane
4.Methanol
5.Pentane
6.Ethanol
7.Acetone
8.Isopropyl alcohol
9.n-Hexane
Peaks
10.Chloroform
11.Benzene
12.Heptane
13.Toluene
14.Ethylbenzene
15.m-Xylene
16.p-Xylene
17.o-Xylene
18.Dimethyl sulfoxide
image
GC_FS0516
ColumnRxi-624Sil MS, 30 m, 0.25 mm ID, 1.40 µm (cat.# 13868)
Sample
Diluent:Dimethyl sulfoxide
Conc.:10 ppm (concentration reflects 100 mg of sample at 50 ppm cutoff)
Injectionheadspace-loop split (split ratio 10:1)
Liner:Premium 1 mm straight (cat.# 23333.1)
Headspace-Loop
Inj. Port Temp.:200 °C
Instrument:Tekmar HT3
Inj. Time:1.00 min
Transfer Line Temp.:160 °C
Valve Oven Temp.:160 °C
Sample Temp.:140 °C
Platen temp equil. time:1.0 min
Sample Equil. Time:10.00 min
Vial Pressure:20 psi
Pressurize Time:5.0 min
Pressure Equilibration Time:0.20 min
Loop Pressure:15 psi
Loop Fill Time:2.0 min
Oven
Oven Temp.:35 °C (hold 1.50 min) to 175 °C at 40 °C/min to 225 °C at 30 °C/min
Carrier GasHe, constant flow
Linear Velocity:80 cm/sec
DetectorFID @ 250 °C
Make-up Gas Flow Rate:45 mL/min
Hydrogen flow:40 mL/min
Air flow:450 mL/min
Data Rate:20 Hz
InstrumentAgilent/HP6890 GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

residual solvent

,

butane

,

solvents

,

624sil MS

,

cannabis

,

,

marijuana

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us