share:

This chromatogram may contain analytes that are synonyms of your search terms.

Residual Solvents in Cannabis Concentrates on Rxi-624Sil MS by Headspace–Full Evaporation Technique (HS-FET)

PeakstR (min)
1.Isobutane0.910
2.Butane0.996
3.Methanol1.112
4.Pentane1.508
5.Ethanol1.545
6.Acetone1.795
7.Isopropanol1.891
8.n-Hexane2.412
9.Chloroform2.962
10.Benzene3.213
11.Heptane3.363
12.Toluene4.135
image
GC_FS0517
ColumnRxi-624Sil MS, 30 m, 0.25 mm ID, 1.40 µm (cat.# 13868)
SampleResidual solvent mix
Diluent:Dimethyl sulfoxide (DMSO)
Conc.:500 ppm (For the HS-FET technique, 10 µL of a 1,000 µg/mL standard was placed into a 20 mL headspace vial to represent a 500 ppm sample concentration, assuming a 20 mg sample.)
Injectionheadspace-loop split (split ratio 10:1)
Liner:Premium 1.0 mm ID straight inlet liner (cat.# 23333.1)
Headspace-Loop
Inj. Port Temp.:250 °C
Instrument:Tekmar HT3
Inj. Time:1.0 min
Transfer Line Temp.:160 °C
Valve Oven Temp.:160 °C
Needle Temp.:140 °C
Sample Temp.:140 °C
Platen temp equil. time:1.0 min
Sample Equil. Time:30.0 min
Vial Pressure:20 psi
Pressurize Time:5.0 min
Loop Pressure:15 psi
Loop Fill Time:2.0 min
Oven
Oven Temp.:35 °C (hold 1.5 min) to 300 °C at 30 °C/min (hold 2.0 min)
Carrier GasHe, constant flow
Linear Velocity:80 cm/sec
DetectorFID @ 250 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
Data Rate:20 Hz
InstrumentAgilent/HP6890 GC
NotesThe FID temperature in this chromatogram was set too low in error. FID temperature should be set to 320°C for residual solvent analysis. This will not affect peak shape or elution profile, but will ensure terpenes do not condense on FID jet.

The butane used for standard preparation was a mixture of butane and isobutane in an unknown ratio. The concentrations of butane and isobutane should be considered approximate, but do not exceed 500 ppm for either component.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

process solvents

,

Marjuanna

,

624Sil

,

concentrates

,

residual solvents

,

,

Cannabis

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us