share:

This chromatogram may contain analytes that are synonyms of your search terms.

Terpene Profile of Citra Hops on Rxi-624Sil MS by FET-HS-GC

PeakstR (min)
1.α-Pinene7.39
2.β-Myrcene8.02
3.α-Phellandrene8.40
4.δ 3-Carene8.44
5.Ocimene8.61
6.Limonene8.71
PeakstR (min)
7.p-Cymene8.75
8.Linalool9.87
9.dl-Menthol11.08
10.Geraniol11.82
11.Citral-212.24
12.β-Caryophyllene13.83
13.α-Humulene14.21
image
GC_FS0525
ColumnRxi-624Sil MS, 30 m, 0.25 mm ID, 1.40 µm (cat.# 13868)
Sample
Conc.: 10 mg of ground Citra hops
Injectionheadspace-loop split (split ratio 10:1)
Liner:Premium 1.0 mm ID straight inlet liner (cat.# 23333.1)
Headspace-Loop
Inj. Port Temp.:250 °C
Instrument:Tekmar HT-3
Inj. Time:1.0 min
Transfer Line Temp.:160 °C
Valve Oven Temp.:160 °C
Needle Temp.:140 °C
Standby flow rate:50 mL/min
Sample Temp.:140 °C
Platen temp equil. time:1.0 min
Sample Equil. Time:30.0 min
Vial Pressure:20 psi
Pressurize Time:5.0 min
Pressure Equilibration Time:0.20 min
Loop Pressure:15 psi
Loop Fill Time:2.0 min
Oven
Oven Temp.:60 °C (hold 0.10 min) to 300 °C at 12.50 °C/min (hold 3.0 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
Linear Velocity:33 cm/sec
DetectorFID @ 320 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
Data Rate:20 Hz
InstrumentAgilent/HP6890 GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

624Sil

,

marihuana

,

cannais

,

terpenes

,

headspace

,

624

,

marijuana

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us