share:

This chromatogram may contain analytes that are synonyms of your search terms.

Benz[a]anthracene, Triphenylene, and Chrysene on Rxi-PAH After Repeated Matrix Injections

Peaks
1.Benz[a]anthracene
2.Triphenylene
3.Chrysene
image
GC_FS0543
ColumnRxi-PAH, 40 m, 0.18 mm ID, 0.07 µm (cat.# 49316)
SampleCustom PAH standard
Diluent:Toluene
Injection
Inj. Vol.:0.5 µL pulsed splitless
Liner:Premium 2 mm single taper w/wool (cat.# 23316.1)
Inj. Temp.:275 °C
Pulse Pressure:80 psi (551.6kPa)
Pulse Time:0.6 min
Purge Flow:40 mL/min
Oven
Oven Temp.:110 °C (hold 1 min) to 210 °C at 35 °C/min to 260 °C at 3 °C/min to 350 °C at 10 °C/min (hold 4.5 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
DetectorFID @ 360 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
InstrumentAgilent 7890A GC
NotesThe experiment was carried out as follows:

- Oysters and paprika were extracted using QuEChERS methodology.
- dSPE cleanup was performed on oyster extracts.
- Cartridge cleanup was performed on paprika extracts.
- Oyster and paprika extracts were injected, interspersed with solvent standard, to monitor performance.

1 m was trimmed from the column and the liner was changed during the course of the experiment.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

EFSA

,

triphenylene

,

chrysene

,

PAH

,

,

Rxi-PAH

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us