Benz[a]anthracene, Triphenylene, and Chrysene on Rxi-PAH After Repeated Matrix Injections
Peaks | |
---|---|
1. | Benz[a]anthracene![]() |
2. | Triphenylene![]() |
3. | Chrysene![]() |

GC_FS0543
Column | Rxi-PAH, 40 m, 0.18 mm ID, 0.07 µm (cat.# 49316) |
Sample | Custom PAH standard |
Diluent: | Toluene |
Injection | |
Inj. Vol.: | 0.5 µL pulsed splitless |
Liner: | Premium 2 mm single taper w/wool (cat.# 23316.1) |
Inj. Temp.: | 275 °C |
Pulse Pressure: | 80 psi (551.6kPa) |
Pulse Time: | 0.6 min |
Purge Flow: | 40 mL/min |
Oven | |
Oven Temp.: | 110 °C (hold 1 min) to 210 °C at 35 °C/min to 260 °C at 3 °C/min to 350 °C at 10 °C/min (hold 4.5 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.4 mL/min |
Detector | FID @ 360 °C |
Make-up Gas Flow Rate: | 45 mL/min |
Make-up Gas Type: | N2 |
Hydrogen flow: | 40 mL/min |
Air flow: | 450 mL/min |
Instrument | Agilent 7890A GC |
Notes | The experiment was carried out as follows: - Oysters and paprika were extracted using QuEChERS methodology. - dSPE cleanup was performed on oyster extracts. - Cartridge cleanup was performed on paprika extracts. - Oyster and paprika extracts were injected, interspersed with solvent standard, to monitor performance. 1 m was trimmed from the column and the liner was changed during the course of the experiment. |