share:

3-Monochloropropane-1,2-Diol (3-MCPD) Esters in Olive Oil on Rxi-17Sil MS

PeakstR (min)Conc.
(µg/g)
1.3-MCPD-d5 PBA derivative5.9371.81
2.3-MCPD PBA derivative5.9772.94
image
GC_FS0566
ColumnRxi-17Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 14123)
SampleExtracted olive oil
Diluent:Isooctane
Injection
Inj. Vol.:1 µL split (split ratio 10:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:280 °C
Oven
Oven Temp.:120 °C (hold 0.5 min) to 180 °C at 12 °C/min to 330 °C at 25 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
DetectorMS
Mode:SIM
SIM Program:147, 150, 196, 201 m/z, 50 ms dwell
Transfer Line Temp.:320 °C
Analyzer Type:Quadrupole
Source Temp.:230 °C
Quad Temp.:150 °C
InstrumentAgilent 7890A GC & 5975C MSD
NotesThe olive oil sample was spiked with 3-MCPD-d5 ester and prepared according to AOCS method Cd 29c-13.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

3-MCPD

,

Olive oil

,

glycidyl ester

,

3-MCPD ester

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us