share:

Bleed Comparison of Rxi-624Sil MS and VF-624ms

Peaks
1.Fluorobenzene
image
GC_GN1147
ColumnRxi-624Sil MS (see notes), 30 m, 0.25 mm ID, 1.4 µm (cat.# 13868)
SampleFluorobenzene (cat.# 30030)
Diluent:methanol
Conc.:200 µg/mL
Injection
Inj. Vol.:1 µL split (split ratio 20:1)
Liner:4mm Split Liner with Wool (cat.# 20781)
Inj. Temp.:220 °C
Oven
Oven Temp.:40 °C (hold 5 min) to 60 °C at 20 °C/min (hold 5 min) to 120 °C at 20 °C/min (hold 5 min) to 200 °C at 20 °C/min (hold 10 min) to 260 °C at 20 °C/min (hold 10 min) to 300 °C at 20 °C/min (hold 20 min)
Carrier GasHe, constant flow
Linear Velocity:40 cm/sec
DetectorFID @ 250 °C
InstrumentAgilent/HP6890 GC
NotesColumns are of equivalent dimensions and were tested after equivalent conditioning.

RELATED CHROMATOGRAM LIBRARY SEARCHES

624

,

bleed

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us