share:

This chromatogram may contain analytes that are synonyms of your search terms.

QuEChERS Extracts of Tylenol Spiked with Anisoles and TBA on Rxi-5Sil MS

image
GC_GN1152
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
SampleQuEChERS extracts of Tylenol spiked with TBA and anisoles
Diluent:Acetonitrile
Injection
Inj. Vol.:1 µL splitless (hold 1 min)
Liner:Gooseneck Splitless (4mm) w/Wool (cat.# 22405)
Inj. Temp.:250 °C
Purge Flow:40 mL/min
Oven
Oven Temp.:90 °C (hold 1 min) to 310 °C at 4 °C/min (hold 2 min)
Carrier GasHe, constant flow
Flow Rate:1.2 mL/min
Linear Velocity:40 cm/sec
DetectorTOFMS
Transfer Line Temp.:290 °C
Analyzer Type:TOF
Source Temp.:225 °C
Electron Energy:70 eV
Mass Defect:-20 mu/100 u
Solvent Delay Time:2 min
Tune Type:PFTBA
Ionization Mode:EI
Acquisition Range:45 to 550 amu
Spectral Acquisition Rate:5 spectra/sec
InstrumentLECO Pegasus 4D GCxGC-TOFMS
NotesTotal ion chromatogram (pink) of QuEChERS extracts of Tylenol spiked with anisoles and TBA, and extracted ion chromatogram (blue, 265x50) of pentachloroanisole, which was located by an automated peak find routine underneath the large acetaminophen peak. No dSPE was performed on this sample; see Applications Note GNAN1263 for complete sample fortification and extraction details.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

tylenol

,

quechers

,

tribromoanisole

,

pentachloroanisole

,

Rxi-5SilMS

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us